Jaszczak SPECT Phantom
Request a QuoteECT Phantom for PET and SPECT
The Jaszczak SPECT Phantom provides consistent performance information for any SPECT or PET system. Multiple performance characteristics of camera-based SPECT systems are evaluated from a single scan of the phantom.
On-axis and off-axis transverse line spread function may be easily measured without removing the cover plate. Measurements of full-width-half (or tenth) maximum can be readily determined, either in air or in water.
The Phantom is used for
- System performance valuation of: Collimator, Artifacts, Calibration, and Reconstruction Parameters
- Acceptance testing
- Routine quality, assurance and control
- Evaluation of center-of-rotation error
- Evaluation of non-uniformity artifact
- Evaluation of changes of radius-of-rotation on spatial resolution
- Evaluation of reconstruction filters on spatial resolution
- Evaluation of attenuation and scatter compensation
- Single slice volume sensitivity
- Total system volume sensitivity
- Lesion detectability
Specifications
- Cylinder Interior Dimensions: 8.5″ dia x 7.32″ h (21.6 x 18.6 cm)
- Cylinder Wall Thickness: 0.125″ (3.2 mm)
- Volume: 6.9 LVolume With Inserts: 6.1 L
- Cold Rod Insert Height: 3.46″ h (8.8 cm)
- Height of Spheres From Base Plate: 5″ h (12.7 cm)
Available in 2 Models
Standard SPECT Phantom
FOR LOWER RESOLUTION CAMERAS
Item # 043-762
Cold Rod Dimensions
6.4 mm, 7.9 mm, 9.5 mm, 11.1 mm, 12.7 mm, 16.0 mm
Solid Sphere Diameters
12.7 mm, 15.9 mm, 19.1 mm, 25.4 mm, 31.8 mm, 38 mm
Deluxe SPECT Phantom
FOR HIGH RESOLUTION CAMERAS
Item # 043-750
Cold Rod Dimensions
4.8 mm, 6.4 mm, 7.9 mm, 9.5 mm, 11.1 mm, 12.7 mm
Solid Sphere Diameters
9.5mm, 12.7 mm, 15.9 mm, 19.1 mm, 25.4 mm, 31.8 mm